advertisement

WGA Rescources

Abstract #51676 Published in IGR 14-4

Ahmed glaucoma valve via pars plana access : Long-term results of implantation for therapy refractive glaucoma

Mazinani B; Schwarzer H; Willkomm A; Weinberger A; Plange N; Walter P; Rössler G
Ophthalmologe 2013; 110: 537-542


BACKGROUND: AIM: of this study is the presentation of long-term results regarding the effectiveness and safety of the implantation of Ahmed glaucoma valve (AGV) devices using a pars plana approach after vitrectomy in the treatment of therapy refractive glaucoma. METHODS: The implantation of AGV devices using a pars plana approach after vitrectomy was performed in 27 eyes of 22 patients with neovascular glaucoma (n = 7 patients), uveitis (n = 6), complex juvenile secondary glaucoma (n = 4), primary open angle glaucoma (n = 3), pseudoexfoliation glaucoma (n = 1) and nanophthalmus (n = 1). Intraocular pressure, visual acuity and subsequent complications were documented. The mean follow-up period was 23.6 months. RESULTS: Intraocular pressure was significantly reduced from 30.2 mmHg preoperatively to 13 mmHg after a follow-up of 36 months. The mean visual acuity did not change significantly, five eyes required a subsequent explantation of which three were due to AGV-related complications. CONCLUSIONS: The implantation of AGV devices using a pars plana approach permits an effective and permanent reduction of intraocular pressure (IOP) even in severe therapy refractive secondary glaucoma; however, the risk of serious sight-threatening complications has to be taken into account.

Universitätsaugenklinik Aachen, Pauwelsstr. 30, 52074, Aachen, Deutschland, bmazinani@gmx.de.

Full article

Classification:

12.8.2 With tube implant or other drainage devices (Part of: 12 Surgical treatment > 12.8 Filtering surgery)



Issue 14-4

Change Issue


advertisement

Oculus