advertisement
BACKGROUND: Analysis of the comparison of flicker perimetry with the prototype Pulsar and the new Octopus 600. METHODS: Both eyes of 20 patients with primary open-angle glaucoma were studied with 30° visual field flimmer perimetry (dynamic strategy) at the perimeters Pulsar and Octopus 600 (Haag-Streit). The evaluation was based on the mean defect (MD = mean deviation) and the defect depth (square root of lost variance = sLV). RESULTS: In the t test for paired samples, the Pulsar perimetry showed significantly higher values than the Octopus 600 perimetry: right eye MD 4.8 ± 3.6 src vs. 1.7 ± 2.9 src, p = 0.005; left eye MD 3.9 ± 3.6 src v. 1.4 ± 2.8 src, p = 0.018; both eyes MD 4.35 ± 3.62 src vs. 1.55 ± 2.80 src, p = 0.002. The sLV values with the Pulsar perimetry were significantly higher than the values with the Octopus 600 perimetry: right eye sLV 3.6 ± 1.6 vs. 2.3 ± 1.3 src, p = 0.006; left eye sLV 3.2 ± 0.8 vs. 2.0 ± 0.8 src, p < 0.0001; both eyes sLV 3.37 ± 1.28 src vs. 2.12 ± 1.05 src, p < 0.0001. The significances even persisted after the Bonferroni-Holm correction. CONCLUSION: The investigation results of flicker perimetry are not comparable, because there is a significant difference in MD and sLV. This mainly due to different light intensities, background brightness and varying colour scalings of the perimetric device.
Augenheilkunde, Augenklinik am Wittenbergplatz, Berlin.
Full article6.6.3 Special methods (e.g. color, contrast, SWAP etc.) (Part of: 6 Clinical examination methods > 6.6 Visual field examination and other visual function tests)