advertisement

WGA Rescources

Abstract #51831 Published in IGR 14-4

Comparison of Goldmann applanation and dynamic contour tonometry in a population of Mexican open-angle glaucoma patients

Jimenez-Roman J; Gil-Carrasco F; Martinez A; Albis-Donado O; De la Torre-Tovar JD
International Ophthalmology 2013; 33: 221-225


To compare intraocular pressure (IOP) measurements obtained with Goldmann applanation (GAT) and dynamic contour tonometry (DCT) in a Mexican population. 40 glaucoma patients were included in this cross-sectional observational cohort study. IOP measurements were performed in the following order: DCT, ultrasonic pachymetry and GAT, with a 5-minute difference between each measurement, between 8 am and 2 pm. Only DCT measurements of good quality (Q ≤ 3) were accepted. GAT measurements were made three times with the same Goldmann tonometer, previously checked for calibration errors, and the mean was used for statistical purposes. The IOP (mean [standard deviation], 95 % confidence interval [CI]) measured with the Goldmann tonometer (13.2 [2.4], 12.4-14.0 mmHg) was significantly lower than that obtained with the DCT (18.4 [3.3], 17.0-19.2 mmHg), p < 0.0001. Pearson's correlation coefficients between CCT and IOP measured with GAT and DCT were (r = 0.24, 95 % CI = 0.07-0.52, p = 0.133) and (r = 0.13, 95 % CI = -0.19 to 0.43, p = 0.412), respectively. The concordance correlation coefficient between GAT and DCT was r (c) = 0.3, 95 % CI = 0.17-0.41). DCT seems to overestimate the IOP as compared to GAT. Additionally, although there was a good correlation between the IOP measurements assessed with either GAT or DCT, the agreement was poor.

Association for Preventing Blindness in Mexico (Dr. Luis Sanchez Bulnes Hospital), Mexico City, Mexico, jimenez_roman@yahoo.com.

Full article

Classification:

6.1.1 Devices, techniques (Part of: 6 Clinical examination methods > 6.1 Intraocular pressure measurement; factors affecting IOP)



Issue 14-4

Change Issue


advertisement

Oculus