advertisement

Topcon

Abstract #8707 Published in IGR 5-1

Reproducibility of measurements of the retinal nerve fibre layer thickness: comparison of OCT with NFA and HRT techniques

Klemm M; Rumberger E; Walter A; Richard G
Ophthalmologe 2002; 99: 345-351


BACKGROUND: Glaucoma is associated with a progressive loss of retinal ganglion cell axons and the evaluation of the thickness of the retinal nerve fiber layer (RNFL) is important for early diagnosis and follow-up. Results of RNFL thickness measurements may vary according to the devices used. Therefore, the present study analyzed and compared the reproducibility of data obtained with three different instruments. METHODS: RNFL thickness was determined using a circular identification line around the optic nerve head by optical coherence tomography (OCT), a Heidelberg Retina Tomograph (HRT), and a Nerve Fiber Analyzer (NFA). Each eye was tested six times. The values for RNFL thickness were calculated for the superior, inferior, nasal and temporal areas. A one-way analysis of variance with random effects was chosen to estimate the variance components. The intraclass correlation coefficient (ICC) as a measure of reliability and the coefficient of variation (CV%) as a measure of precision were determined to estimate the reproducibility of RNFL thickness measurements. RESULTS: With all three instruments, the reproducibility of measurements between the superior, inferior, temporal, and nasal areas did not differ significantly, suggesting that reproducibility does not depend on the thickness of the RNFL. Data obtained by NFA showed the highest ICC values (0.9) and lowest CV% values (7.0), followed by those obtained by HRT (reliability ICC 0.8; precision CV% 12). A markedly lower degree of reproducibility was found for the OCT compared to NFA and HRT (reliability ICC 0.5; precision CV% 15). CONCLUSIONS: The comparison of RNFL thickness measurements with OCT, NFA, and HRT showed that the NFA yielded the most reliable results followed by HRT.LA: German

Dr. M. Klemm, Universitatsaugenklinik Hamburg, Martinistrasse 52, 20246 Hamburg, Germany


Classification:

2.13 Retina and retinal nerve fibre layer (Part of: 2 Anatomical structures in glaucoma)
6.9.1 Laser scanning (Part of: 6 Clinical examination methods > 6.9 Computerized image analysis)
6.9.2 Optical coherence tomography (Part of: 6 Clinical examination methods > 6.9 Computerized image analysis)



Issue 5-1

Change Issue


advertisement

Nidek